SImA analyses results are not visible in Spotfire

Last modified by Martin SZOTKOWSKI on 2025/02/10 08:06

Problem: In Spotfire, there are not accessible all results from SImA but only the latest one.

SImA:image-2025-2-10_9-0-38.png

Spotfire:image-2025-2-10_9-1-15.png

Root cause: Results are stored in two formats: xml (used in SImA internally) and csv (used in import to Spotfire). When data were transferred from Columbus, all the xml files were transferred so they are visible in SImA, but only the latest csv file was transferred, so import to Spotfire is possible only for latest result.

Solution: Run the analyses once again, then the data are stored in both formats and are accessible in Spotfire.


Revvity support ticket number: 298858